Fluke Speaker Rev. 4 Manuale Utente Pagina 66

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 118
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 65
701/702
Users Manual
56
and connect the calibrator to the circuit to be
tested. Figure 20 shows an example.
2. If you want to automatically stop ramping if a trip
condition is detected, connect a voltage trip
circuit to the V MEAS jacks or a continuity trip
circuit to the mA
RTD MEAS jacks. (Continuity
detection is not available when sourcing current.)
3. If necessary, press M for SOURCE mode.
4. Set the calibrator for the desired source value as
previously described.
5. If you want to ramp the output in % of scale, set
% of scale as described previously under
“Sourcing in Percent of Scale.”
6.
Press the More Choices softkey.
7.
Press the Ramp softkey. The display changes to
the following:
Vedere la pagina 65
1 2 ... 61 62 63 64 65 66 67 68 69 70 71 ... 117 118

Commenti su questo manuale

Nessun commento