Fluke 8842a Manuale Utente Pagina 113

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 112
8842A
Instruction Manual
5-6
5-5. VDC Scaling
Scaling is performed in the VDC function by two precision resistors networks (Z301 and
Z302). These components are configured by relay K301, switching transistor Q311, and
quad analog switches U302A and U301B to provide the correct scaling for each range.
Voltage follower U306 provides high input impedance for the 20V dc range. A simplified
schematic and a switch state table for the VDC function are shown in Figure 5-2.
f5-02.wmf
Figure 5-2. DC Scaling (VDC and mA DC)
Vedere la pagina 112
1 2 ... 108 109 110 111 112 113 114 115 116 117 118 ... 268 269

Commenti su questo manuale

Nessun commento